50kVA/100kv Pd Free Test System

Product Details
Customization: Available
After-sales Service: 1 Year
Warranty: 1 Year

360° Virtual Tour

Diamond Member Since 2005

Suppliers with verified business licenses

Audited Supplier

Audited by an independent third-party inspection agency

Number of Employees
12
Year of Establishment
2004-08-25
  • 50kVA/100kv Pd Free Test System
  • 50kVA/100kv Pd Free Test System
  • 50kVA/100kv Pd Free Test System
  • 50kVA/100kv Pd Free Test System
  • 50kVA/100kv Pd Free Test System
  • 50kVA/100kv Pd Free Test System
Find Similar Products

Basic Info.

Model NO.
ZH
Type
Pd Free Test System
Accuracy Grade
0.5
Display
Digital
Control
Computer Control
Power Source
AC220V
Rated Capacity
50kVA
Output Voltage
0-400V, 0-100kv Adjustable
Output Voltage Frequency
50-300Hz
Phase Number
Single
Short-Circuit Impedance
<5%
Noise
<65 dB
No-Load Current
<10%
Pd Amount at Rated Voltage
<5PC
Trademark
zhenhuan
Origin
China
Production Capacity
5000PCS/Year

Product Description

50kVA/100kV  PD Free Test System

I DescriptionThis machine is suitable for power frequency withstand voltage and partial discharge, frequency doubling induction withstand voltage and partial discharge test of high-voltage voltage transformers and current transformers of 33 KV and below.
II Main configuration
Item Model Qty. Remark
Automatic power frequency and frequency doubling PDF test system 50/100 1 The system comes with square wave generator and impedance unit.
PDF test transformer 50KVA/100KV 1
Digital Partial Discharge Tester 124 1
PDF frequency conversion power supply 50KVA 1
Coupling capacitor 100KV 1
Protection resistor 100KV-5KΩ 1


50/100 automatic power frequency and frequency multiplication PDF test system diagram

III Main technical data
  1. Automatic power frequency and frequency multiplication PDF test system diagram
Rated capacity 50kVA
Input voltage 3P, three-wire+ground, 380V±10%, 50Hz/60Hz±5Hz
Output voltage 0-400V, 0-100kV adjustable
Output voltage frequency 50-300Hz
Control Automatic
Phase number Single
Short-circuit impedance <5%
Noise <65 dB
No-load current <10%
PD amount at rated voltage <5pC

  1. Transformer Insulation Level
Transformer winding Rated short-term power frequency withstand voltage kV
 (root mean square)
HV winding Power frequency 100 (5 min)
LV winding 3 (1 min)
 
  1. Automatic PDF control table
Rated voltage AC 380V
Capacitance 50KVA
Input voltage 0-400V
Input frequency 50-300Hz
Voltage accuracy ±0.5
Current accuracy ±0.5
Operation method Automatic, manual
Data preservation 1000 group
Password Multi-level permission settings
Weight 10kg
Six protections
Low voltage protection Low current protection
High voltage protection High current protection
Over voltage protection High Voltage Flash-over Protection
Software operation interface
Operation information prompt During automatic test, 5-stage test can be set
Generate reports automatically Protection status prompt
Voltage and time graph Primary circuit working status display
Power frequency withstand voltage and induction withstand voltage can be switched freely
 
  1. Digital Partial Discharge Tester
  1. Capacitance range of testable products: 6PF--250μF
  2. Detection sensitivity
  3. Amplifier Band
NO. Tuning capacitor Unit Sensitivity(pC)(asymmetric circuit)
1 6-25-100 pF 0.02
2 25-100-400 pF 0.04
3 100-400-1500 pF 0.06
4 400-1500-6000 pF 0.1
5 1500-6000-25000 pF 0.2
6 0.006-0.025-0.1 pF 0.3
7 0.025-0.1-0.4 pF 0.5
8 0.1-0.4-1.5 pF 1
9 0.4-1.5-6.0 pF 1.5
10 1.5-6.0-25 pF 2.5
11 6.0-25-60 pF 5
12 25-60-250 pF 10
7R Resistor   0.5
  1. Low-end: 10KHZ, 20KHZ, 40KHZ(optional)
  2. High-end: 80KHZ, 200KHZ, 300KHZ(optional)
 
  1. Amplifier Gain Adjust
Coarse adjustment six gears, gain between gears 20±1 db; fine adjustment range ≥ 20db
 
  1. Time Window:
1. Window width: adjustable range 15°~150°;
2. Window position: Each window can be rotated 0°~170°;
3. The two time windows can be opened separately or simultaneously.
 
  1. Discharge meter:
Digital meter head: Displayed by 3½ LED digital meter
0-100.0 error <±5% (in full scale)
 
  1. Elliptical time base:
Frequency 50HZ, 100HZ, 150HZ, 200HZ, 400HZ.
Elliptical rotation: 150° rotation is possible with 30° as the first gear.
Display mode: ellipse - straight line - sine.
 
  1. Test voltmeter:
Range: 100KV (expandable)
Display: 3½ digital voltmeter indication
Accuracy: better than ±5% (on full scale)
 
  1. It has the function of zero mark, which is beneficial to the identification of partial discharge
  2.  Dimension: 500 x 500 x 210 mm, weight: 16 kg
 
  1. PDF test transformer
  1. Type: oil immersed
  2. Phase number: single
  3. Rated capacity: 50kVA
  4. Input rated voltage: 400V
Input rated current: 125A
  1. Output voltage: 100kV
Output rated current: 500mA
  1. No-load current: <5%
  2. Waveform distortion rate: ≤3%
  3. Measuring winding voltage: 100V Measuring voltage ratio: 1:1000
  4. Over-current capability: Under the condition of 150% (0.75) rated current for 300S, the transformer winding of the over-current test will not cause thermal damage and winding deformation
  5. Over-voltage capability: 110% UH, over-voltage under 60S will not cause insulation damage to the test transformer, and the waveform distortion rate ≤5%
  6. Insulation level: low-voltage end power frequency resistance 5kV/min
  7. Dielectric loss: ≤0.5%
  8. Partial discharge: less than 5PC at 100% UH, less than 3PC at 80% UH
 
  1. PDF frequency conversion power supply
  1. Rated power: 50kVA
  2. Input voltage: 380V±10%,50Hz/60Hz±5Hz
  3. Output voltage: 0-400V
  4. Output voltage frequency: 45-300Hz(adjustable)
 
  1. PD Free divider (also coupling capacitor)
  1. Capacity: 700pF
  2. Measure accuracy: 1%
  3. Max. Measure voltage: 100KV
  4. Manufactured with epoxy glass fiber tube, coated with hydrophobic insulating paint
  5. All high-voltage components are treated with special technology, and there is no partial discharge phenomenon
 
  1. PDF protection resistor
  1. Model No. 100KV-5
  2. Rated frequency: 60HZ
  3. Rated voltage: 100KV
  4. Rated current: 0.1A
  5. Nominal resistance: 5KΩ
  6. Heat resistance grade: F grade
  7. Partial discharge: <3pC under 100% rated voltage, <2pC under 80% rated voltage
  8. Temperature rise: under 100% rated current, the surface temperature of the resistance in continuous operation is less than 80K
  9. Running time: same as transformer
IV Feature
  1. Pressure regulating control unit
  1. Industrial embedded PC, using high-performance hardware and software control; touch screen operation.
  2. The console adopts automatic control and manual control, and the two functions are managed independently and hierarchically without interfering with each other;
  3. All test parameters of this test system can be arbitrarily set according to requirements, the test speed is smooth, and the displayed data is clear and intuitive;
  4. All test parameters can be set in advance, which can achieve a series of functions of accurate memory, storage and reset of high-voltage voltage discharge breakdown voltage; automatic scanning time setting, automatic storage of detailed records of time, voltage, current, etc. during the electrical aging process, which is convenient for users to view ;
  5. The test system is equipped with a cumulative timing function, and can independently start, pause and clear;
  6. The combination of hardware and software anti-jamming technology has stable performance and strong anti-jamming.
  7. It has power frequency/frequency doubling output capability, and can be used for both power frequency PDF withstand voltage test and frequency doubling PDF withstand voltage test. It can also be used as a frequency multiplier PDF power supply.
  1. PD measurement unit
  1. After the partial discharge signal is digitally processed, it is displayed on the LCD.
  2. Free choice of ellipse, straight line, sine display mode
  3. Adopt digital windowing technology to avoid the influence of interference on measurement.
  4. Window opening in any phase, single window, double window optional, ellipse 150 degree rotation.
  5. The current discharge map can be frozen and saved, and the partial discharge pulse can be observed and analyzed statically.
50kVA/100kv Pd Free Test System50kVA/100kv Pd Free Test System50kVA/100kv Pd Free Test System50kVA/100kv Pd Free Test System